5.9. CALIBRATION QUALITY ANALYSIS

FUNCTION
NOX SLOPE
NO SLOPE
NOX OFFS
NO OFFS
07270D DCN7141
After completing one of the calibration procedures described above, it is important to
evaluate the analyzer's calibration SLOPE and OFFSET parameters. These values
describe the linear response curve of the analyzer, separately for NO and NO
values for these terms, both individually and relative to each other, indicate the quality
of the calibration. To perform this quality evaluation, you will need to record the values
of the following test functions (Section 4.2.1 or Appendix A-3), all of which are
automatically stored in the DAS channel CALDAT for data analysis, documentation
and archival.
NO OFFS
NO SLOPE
NOX OFFS
NOX SLOPE
Make sure that these parameters are within the limits listed in Table 5-5 and frequently
compare them to those values on the Final Test and Checkout Sheet that came attached
to your manual, which should not be significantly different. If they are, refer to the
troubleshooting Section 7.
Table 5-5: Calibration Data Quality Evaluation
MINIMUM VALUE
-0.700
-0.700
-20.0 mV
-20.0 mV
The default DAS configuration records all calibration values in channel CALDAT as
well as all calibration check (zero and span) values in its internal memory. Up to 200
data points are stored for up 4 years of data (on weekly calibration checks) and a lifetime
history of monthly calibrations. Review these data to see if the zero and span responses
change over time. These channels also store the STABIL value (standard deviation of
NO
concentration) to evaluate if the analyzer response has properly leveled off during
X
the calibration procedure. Finally, the CALDAT channel also stores the converter
efficiency for review and documentation.
OPTIMUM VALUE
1.000
1.000
0.0 mV
0.0 mV
Calibration Procedures
. The
X
MAXIMUM VALUE
1.300
1.300
150.0 mV
150.0 mV