Advertisement

Teledyne API - Model T200H/T200M Operation Manual
4.13.6.2. Manual Calibration of Analog Output Configured for Voltage Ranges
Note
See the Electrical
Table 4-24: Voltage Tolerances for Analog Output Calibration
FULL
SCALE
0.1 VDC
1 VDC
5 VDC
10 VDC
07270D DCN7141
For highest accuracy, the voltages of the analog outputs can be manually calibrated.
The menu for manually adjusting the analog output signal level will only appear if
the AUTO-CAL feature is turned off for the channel being adjusted.
Calibration is performed with a voltmeter connected across the output terminals (See
Figure 4-11) and by changing the actual output signal level using the front panel keys in
100, 10 or 1 count increments.
Connections
section for pin
assignments of
Analog Out
connector on the
+DC
rear panel
V OUT +
V OUT -
ANALYZER
Figure 4-11:
Setup for Calibrating Analog Outputs
ZERO
SPAN VOLTAGE
TOLERANCE
±0.0005V
±0.001V
±0.002V
±0.004V
V
Gnd
SPAN
TOLERANCE
90 mV
±0.001V
900 mV
±0.001V
4500 mV
±0.003V
4500 mV
±0.006V
Operating Instructions
V IN +
V IN -
Recording
Device
MINIMUM
ADJUSTMENT
(1 count)
0.02 mV
0.24 mV
1.22 mV
2.44 mV
149