Figure 7-12: Pump Ac Power Jumpers (Jp7) - Teledyne T200H User Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Troubleshooting & Service
7.5.4.1. AC Configuration – Internal Pump (JP7)
Table 7-3: AC Power Configuration for Internal Pumps (JP7)
LINE
LINE
POWER
FREQUENCY
60 HZ
110VAC
115 VAC
50 HZ
60 HZ
220VAC
240 VAC
50 HZ
1
A jumper between pins 5 and 10 may be present on the jumper plug assembly, but has no function on the T200H/M
analyzers.
240
AC power configuration for internal pumps is set using Jumper set JP7 (see Figure 7-4
for the location of JP7).
JUMPER
COLOR
Connects pump pin 3 to 110 / 115 VAC power line
WHITE
Connects pump pin 3 to 110 / 115 VAC power line
Connects pump pin 3 to 110 / 115 VAC power line
1
BLACK
Connects pump pin 3 to 110 / 115 VAC power line
BROWN
1
BLUE
Figure 7-12:
Pump AC Power Jumpers (JP7)
Teledyne API - Model T200H/T200M Operation Manual
FUNCTION
Connects pump pins 2 & 4 to Neutral
Connects pump pins 2 & 4 to Neutral
Connects pump pins 3 and 4 together
Connects pump pin 1 to 220 / 240VAC power line
Connects pump pins 3 and 4 together
Connects pump pin 1 to 220 / 240VAC power line
JUMPER
BETWEEN
PINS
2 to 7
3 to 8
4 to 9
2 to 7
3 to 8
4 to 9
1 to 6
3 to 8
1 to 6
3 to 8
07270D DCN7141