Sample Mode; Test Functions; Table 4-1: Analyzer Operating Modes - Teledyne T200H User Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API - Model T200H/T200M Operation Manual
MODE
SAMPLE
M-P CAL
SETUP X.#
SAMPLE A
ZERO CAL M
ZERO CAL A
ZERO CAL R
LO CAL A
LO CAL R
SPAN CAL M
SPAN CAL A
SPAN CAL R
DIAG
1
Only Appears on units with Z/S valve or IZS options.
2
The revision of the analyzer firmware is displayed following the word SETUP, e.g., SETUP
F.0.

4.2. SAMPLE MODE

4.2.1. TEST FUNCTIONS

07270D DCN7141

Table 4-1: Analyzer Operating Modes

Sampling normally, flashing text indicates adaptive filter is on.
This is the basic calibration mode of the instrument and is activated
by pressing the CAL key.
2
SETUP mode is being used to configure the analyzer. The gas
measurement will continue during this process.
Indicates that unit is in SAMPLE mode and AUTOCAL feature is
activated.
1
Unit is performing ZERO calibration procedure initiated manually by
the user.
1
Unit is performing ZERO calibration procedure initiated automatically
by the AUTOCAL feature.
1
Unit is performing ZERO calibration procedure initiated remotely
through the COM ports or digital control inputs.
Unit is performing LOW SPAN (midpoint) calibration initiated
automatically by the analyzer's AUTOCAL feature.
Unit is performing LOW SPAN (midpoint) calibration initiated remotely
through the COM ports or digital control inputs.
1
Unit is performing SPAN calibration initiated manually by the user.
1
Unit is performing SPAN calibration initiated automatically by the
analyzer's AUTOCAL feature.
1
Unit is performing SPAN calibration initiated remotely through the
COM ports or digital control inputs.
One of the analyzer's diagnostic modes is active (Section 4.13).
The very important CAL mode, which allows calibration of the analyzer in various
ways, is described in detail in Section 5.
This is the analyzer's standard operating mode.
analyzing NO and NO
X
A series of test functions is available at the front panel while the analyzer is in SAMPLE
mode. These parameters provide information about the present operating status of the
instrument and are useful during troubleshooting (Section 7). They can also be recorded
in one of the DAS channels (Section 4.7) for data analysis or output on one of the
configurable analog outputs.
EXPLANATION
and calculating NO
concentrations.
2
Operating Instructions
In this mode, the instrument is
71