Teledyne T200H User Manual page 64

Nitrogen oxides analyzer
Table of Contents

Advertisement

Getting Started
STEP 3 – Set NO
and NO span gas concentrations:
x
The NO
& NO span concentration
X
values automatically default to
80.0 Conc.
If this is not the the concentration of
the span gas being used, toggle
these buttons to set the correct
concentration of the NO
calibration gases.
62
Set the expected NO and NO
concentration values likely to be encountered in this application. The default factory
setting is 100 ppm. If one of the configurable analog outputs is to be set to transmit
concentration values, use 80% of the reporting range set for that output (see Section
4.13.4)
If you supply NO span gas to the analyzer as well as NO
and NO
span gas concentrations need to be identical.
x
SAMPLE
A1:NXCNC1=100PPM
< TST TST >
SAMPLE
NOX
O2
SAMPLE
LOW
HIGH
M-P CAL
A1:NXCNC1 =100PPM
<TST TST> ZERO SPAN
M-P CAL
CONCENTRATION MENU
NOX
NO
CONV
M-P CAL
NOX SPAN CONC:80.0 Conc
0
0
and NO
X
Teledyne API - Model T200H/T200M Operation Manual
span gas concentration. These should be 80% of range of
X
NOX=XXX.X
CAL
SETUP
GAS TO CAL:NOX
ENTR
RANGE TO CAL:LOW
ENTR
NOX=X.XXX
CONC
8
0
.0
ENTR EXIT
, the values for expected NO
x
EXIT
EXIT
EXIT
EXIT
EXIT ignores the new
setting and returns to
the previous display.
ENTR accepts the new
setting and returns to
the
CONCENTRATION
MENU.
If using NO span gas
in addition to NO
X
repeat last step.
07270D DCN7141