Teledyne T200H User Manual page 136

Nitrogen oxides analyzer
Table of Contents

Advertisement

Operating Instructions
4.13.5.2. Analog Output Signal Type and Range Selection
These keys set
the signal level
and type of the
selected
channel
134
To select an output signal type (DC Voltage or current) and level for one output channel
press:
From the
AIO CONFIGURATION SUBMENU
(in the SETUP DIAG menu)
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
DIAG AIO
AOUTS CALIBRATED: NO
SET>
CAL
Continue pressing SET> until you reach the
output to be configured
DIAG AIO
DATA_OUT_3: 5V, NXCNC1, NOCAL
<SET SET>
EDIT
DIAG AIO
DATA_OUT_3: RANGE: 5V
0.1V
1V
5V
10V CURR
Teledyne API - Model T200H/T200M Operation Manual
ENTR
EXIT
EXIT
EXIT
ENTR EXIT
Pressing ENTR records
the new setting and
returns to the previous
menu.
Pressing EXIT ignores the
new setting and returns to
the previous menu.
07270D DCN7141