Remote Das Configuration - Teledyne T200H User Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Operating Instructions

4.7.3. REMOTE DAS CONFIGURATION

Figure 4-4:
Note
94
Editing channels, parameters and triggering events as described in 6.7 is much more
conveniently done in one step through the APICOM remote control program using the
graphical interface shown in Figure 4-4. Refer to Section 4.15 for details on remote
access to the T200H/M analyzer.
APICOM Graphical User Interface for Configuring the DAS
Once a DAS configuration is edited (which can be done offline and without interrupting
DAS data collection), it is conveniently uploaded to the instrument and can be stored on
a computer for later review, alteration or documentation and archival. Refer to the
APICOM manual for details on these procedures.
included in the APICOM installation file, which can be downloaded at
http://www.teledyne-api.com/software/apicom/.
Whereas the editing, adding and deleting of DAS channels and parameters of one
channel through the front-panel touch screen can be done without affecting the
other channels, uploading a DAS configuration script to the analyzer through its
communication ports will erase all data, parameters and channels by replacing
them with the new DAS configuration. It is advised to download and backup all
data and the original DAS configuration before attempting any DAS changes.
Teledyne API - Model T200H/T200M Operation Manual