Calibration Procedures
5.2.5.1. Determining / Updating the NO
STEP ONE:
NO
Gas
2
(High Concentration)
MODEL 701
Zero Gas
Generator
Figure 5-1:
184
The following procedure will cause the Model T200H/M to automatically calculate the
current NO
conversion efficiency.
2
Connect a source of calibrated NO
MODEL T700
Gas Dilution
Calibrator
VENT if not vented
at calibrator
PUMP
Gas Supply Setup for Determination of NO
Teledyne API - Model T200H/T200M Operation Manual
Converter Efficiency
2
span gas as shown below.
2
Source of
SAMPLE GAS
Removed during
calibration
SAMPLE
EXHAUST
VENT
here if input
is pressurized
Instrument
Chassis
Conversion Efficiency
2
07270D DCN7141