TYPE B EQUIPMENT: Equipment providing a particular degree of protec- tion against electric shock, particularly regarding— • • TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments allowable LEAKAGE CURRENT Reliability of the protective earth connection (if present). ECO # 00-0086 P/N TM0021 04/03/00...
Page 2
The class T-7 Micro-Fuel Cell is warranted for one year from the date of shipment from Teledyne. The liability of Teledyne, if any, shall be limited solely to the replacement and repair of the goods, and shall not include shipping costs or other incidental damages as defined in Section 2-715 of the U.S Uniform Commercial Code.
’ S AND OUTINE AINTENANCE ROUBLESHOOTING ECHNICAL PECIFICATIONS FFECT OF RESSURE FFECT OF UMIDITY FFECT OF EMPERATURE ISCREPANCY IN EADING EPAIR ERVICE ENSOR EPLACEMENT ECHNICAL ERVICE ENERAL OTES PARE ARTS TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments ABLE OF ONTENTS ONES UMBER...
ISO 9002 Quality System protocols. When used correctly, this instrument will provide you with many years of trouble free service. The TED 191 Portable Oxygen Monitor is a compact, self-contained unit, which is battery powered by four (4) dry cell batteries (size AA/penlight).
The use of this monitor as a primary or only means of preparing gas mixtures is not advised. The monitor is capable of verifying oxygen concentrations in gas mixtures used in: • Anesthesia • Neonatal Intensive care • Adult Intensive care • Respiratory / Oxygen Therapy TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments PPLICATIONS...
Liquid Crystal Display Displays measured oxygen concentration, low battery condition and provides a visual indication of either a high (+) or low (-) alarm threshold violation. TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments ESCRIPTION OF Alarm Mute Button TELEDYNE...
ANEL Battery Drawer Battery Drawer Sensor Cable Connector Plug TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments ESCRIPTION OF Sensor Cable Connector Plug The Battery Drawer contains 4 x AA (Penlight) batteries required to power the instrument. The Oxygen Sensor Connector Cable is plugged into this “one way”...
ECTION PERATIONS Note : Upon receipt, inspect the entire unit and accompanying accessories for damage or broken or loose parts. If damaged, do not use. Notify the Shipper, and consult Teledyne Electronic Technologies Analytical Instruments 1.1 S To set-up your TED 191 Portable Oxygen Monitor : 1.1.1 Install the Batteries...
Page 9
Care should be taken to ensure that the correct battery polarity is maintained. Caution : Improper installation of the batteries may result in heat damage to the batteries. Re-insert the battery drawer into the monitor casing, and close. TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments PERATIONS...
Page 10
This jack will only fit one way, so do not force it. Note : For sterilization procedures on the Oxygen Cell, see Section 3 on Page 15 below. Caution : Do NOT autoclave the oxygen sensor TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments PERATIONS...
It is not uncommon for the ambient air in hospital wards to be as high as 23%. If the unit fails the 21% calibration reading move the instrument to a location where oxygen contamination is unlikely. TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments PERATIONS...
The magnitude of this fall is related to the level of oxygen concentration and the duration of the exposure to the anaesthetic agent. See 1.7 below for further details. TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments PERATIONS being monitored.
Audible alarm signals are tonally matched to the type of threshold violations i.e. low alarm violations are signalled via a lower pitched audible signal, while high alarm violations are signalled via a higher pitched audible signal. TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments ±...
ISO 7767: 1997 (E) Caution: The TED 191 Portable Oxygen Monitor should not be used in the presence of flammable anaesthetics such as diethal ether or cyclpropane. The T-7 oxygen sensor should not be left in nitrous oxide mixtures any longer than absolutely necessary.
ECTION HREE 3.1 C TED 191 P LEANING THE The instrument itself should be cleaned using a damp cloth. In the event of an excessive build-up of dirt, a mild alcohol solution (isopropyl) may be used to wipe over the instrument, which should be allowed to air-dry afterwards.
Expose instrument to other equipment with high levels of RFI emission. Place the instrument itself in a water vapour saturated environment. Expose the LCD to excessive direct sunlight Dispose of the sensor cell in fire, or place in an incinerator. TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments ’ ’...
A Sensor Cell Replacement Log has been provided for the reference of the users, and is located at the back of this Instruction Manual. The details of the replacement sensor should be entered into this log at the time of changing. TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments ENSOR...
1) Remove the sensor tip from the T-adaptor, unscrew the “flow changes in oxygen diverter” and gently swab any excess moisture off the sensor cell concentration membrane, then flow dry gas over the membrane. TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments ROUBLESHOOTING EMARK : check if functional on batteries.
Page 19
2 – 3% over 3 – 4 hours 2) Replace cable. Alarm continuously activates when 1) The minimum low alarm threshold is 18%. The Teledyne TED 191 displayed value is less will always alarm under these conditions. than 18% TELEDYNE ELECTRONIC TECHNOLOGIES...
MEASUREMENT RANGE 0 to 100% of localised environmental oxygen concentration DISPLAY TYPE Liquid Crystal Display (LCD) SENSOR TYPE Class T-7 (Galvanic) ACCURACY Approximately 2% linear SENSOR LIFE Approximately 12 months in most applications TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments ECHNICAL PECIFICATIONS...
Page 21
Operation of instrument is unaffected by : GENERAL Multiple mounting bracket configurations to suit any mounting requirement. (Custom engineered solutions available on request). TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments ECHNICAL Humidity (water vapour saturated gas); Pressure (up to 100cm H...
This is why it is important to calibrate the TED 191 Portable Oxygen Monitor at regular intervals. It is recommended that the instrument be calibrated prior to each use or every 8 hours.
For example, if an 80% oxygen gas mixture is humidified to saturation at room temperature, the resulting gas mixture will contain only 77.5% oxygen. Your TED 191 Portable Oxygen Monitor accurately measures decreases in the oxygen concentration due to the dilution effects of moisture added to gas mixtures.
ECTION EMPERATURE The Teledyne Electronic Technologies changes in the range of 0 - 40ºC (32 - 106ºF). Since the thermistor that compensates for these changes is located in the rear of the sensor assembly, it is important that gas mixtures, flowing over the front of the sensor, be at room temperature.
LEVEN EADINGS The TED 191 is intended to be used as a secondary oxygen monitor, meaning that it is intended to verify the accuracy of and check the oxygen concentration leaving another oxygen mixing device or primary life support system (i.e a blender or anaesthesia machine). Whenever there is a significant difference in the oxygen readings between the primary and secondary monitors, the discrepancy must be resolved immediately.
ECTION WELVE ERVICE In the event that your TED 191 Portable Oxygen Monitor requires service, the following steps will help to ensure that the repair request is processed promptly : Contact your authorised Teledyne Electronic Technologies or factory for return instructions. Do not ship your monitor without first obtaining authorisation.
Orders for spare parts should include the model number, serial number, and range of the analyzer for which the parts are intended. Orders should be sent to : Teledyne Electronic Technologies Analytical Instruments 16830 Chestnut Street City of Industry, CA 91749-1580...