Manual Calibration With Zero/Span Valves - Teledyne T200H User Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API - Model T200H/T200M Operation Manual

5.5. MANUAL CALIBRATION WITH ZERO/SPAN VALVES

STEP ONE:
MODEL T700
Gas Dilution
Calibrator
MODEL 701
Zero Gas
Generator
Figure 5-4:
07270D DCN7141
Zero and Span calibrations using the Zero/Span Valve option are similar to that
described in Section 5.3, except that:
Zero air and span gas is supplied to the analyzer through the zero gas and span
gas inlets rather than through the sample inlet.
The zero and cal operations are initiated directly and independently with dedicated
keys (CALZ & CALS)
If both available DAS parameters for a specific gas type are being reported via the
instruments analog outputs, e.g. NXCNC1 and NXCNC2, separate calibrations should
be carried out for each parameter.
Use the LOW button when calibrating for NXCNC1
Use the HIGH button when calibrating for NXCNC2.
See Section 4.13.4 for more information on analog output reporting ranges
Connect the sources of zero air and span gas to the respective ports on the rear panel as
shown below.
VENT if not
vented at
calibrator
External Zero
r
Air Scrubbe
Pneumatic Connections–With Zero/Span Valve Option (50)
Source of
SAMPLE Gas
PUMP
Filter
Calibration Procedures
VENT
here if input
is pressurized
Sample
Exhaust
Span Point
Instrument
Zero Air
Chassis
193