Teledyne T200H User Manual page 149

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API - Model T200H/T200M Operation Manual
4.13.6.1. Automatic Analog Output Calibration
Analyzer
automatically
calibrates all
channels for which
AUTO-CAL is turned
ON
If any of the channels
have not been calibrated
ot if at least one channel
has AUTO-CAL turned
OFF, this message will
read NO.
Note
07270D DCN7141
To calibrate the outputs as a group with the AOUTS CALIBRATION command, press:
AIO CONFIGURATION SUBMENU
DIAG
PREV NEXT
DIAG AIO
SET>
DIAG AIO
DIAG AIO
DIAG AIO
DIAG AIO
DIAG AIO
SET> CAL
Manual calibration should be used for the 0.1V range or in cases where the
outputs must be closely matched to the characteristics of the recording device.
From the
(in the SETUP DIAG menu)
ANALOG I/O CONFIGURATION
ENTR
AOUTS CALIBRATED: NO
CAL
AUTO CALIBRATING DATA_OUT_1
AUTO CALIBRATING DATA_OUT_2
NOT AUTO CAL. DATA_OUT_3
AUTO CALIBRATING DATA_OUT_4
AOUTS CALIBRATED: YES
Operating Instructions
EXIT
EXIT
This message
appears when
AUTO-CAL is
Turned OFF for
a channel
EXIT
147