8.3.8. OZONE SUPPLY AIR FILTER

8.3.9. OZONE SCRUBBER

07270D DCN7141
The T200H/M uses ambient air as the supply gas for the O
variety of byproducts. Small amounts of water, ammonia and various sulfur oxides can
combine to create ammonium sulfate, ammonium nitrate, nitric acid and other
compounds.
Whereas sulfates and nitrates can create powdery residues inside the
reaction cell causing sensitivity drift, nitric acid is a very aggressive compound, which
can deteriorate the analyzer's components. In order to remove these chemical byproducts
from the O
gas stream, the output of the O
3
between the generator and the reaction cell.
Any NO
that may be produced in the generator (from reaction of O
X
air) and may cause an artifact in the measurement, is calibrated out through the Auto-zero
functionality, which checks the background signal of the O
Even though ozone is unstable and typically reacts to form O
quite fast enough to ensure that it is completely removed from the exhaust gas stream of
the T200H/M by the time the gas exits the analyzer. Due to the high toxicity and
reactivity of O
, a special catalytic ozone scrubber is used to remove all of the O
3
the reaction cell. Besides its efficient destruction of O
toxic or hazardous gases as it only converts ozone to oxygen.
The O
scrubber is located inside the NO
3
order to utilize residual heat given of by the converter heater. Even though the catalyst is
100% efficient at scrubbing ozone at room temperature, heating it significantly reduces
the necessary residence time (the amount of time the gas must be in contact with the
catalyst) for 100% efficiency and full efficiency can be maintained at higher gas flow
rates. As this is a true catalytic converter, there are no maintenance requirements as
would be required for charcoal-based scrubbers.
A certain amount of fine, black dust may exit the catalyst, particularly if the analyzer is
subjected to sudden pressure drops (for example, when disconnecting the running pump
without letting the analyzer properly and slowly equilibrate to ambient pressure). To
avoid the dust from entering the reaction cell or the pump, the scrubber is equipped with
sintered stainless steel filters of 20 µm pore size on either end and on some models, an
additional dust filter may be attached to the exhaust port.
generator and may produce a
3
generator flows through a special filter
3
stream only once per minute.
3
2
, this catalyst does not produce any
3
converter housing next to the NO
2
Principles of Operation
or O
and N
in the
2
3
2
, the break-down is not
exiting
3
converter in
2
283