Teledyne T200H User Manual page 155

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API - Model T200H/T200M Operation Manual
4.13.6.4. AIN Calibration
07270D DCN7141
This is the sub-menu calibrates the analyzer's A-to-D conversion circuitry.
calibration should only be necessary after major repair such as a replacement of CPU,
motherboard or power supplies.
To perform a AIN CALIBRATION, press:
AIO CONFIGURATION SUBMENU
(in the SETUP DIAG menu)
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
DIAG AIO
AOUTS CALIBRATED: NO
SET>
CAL
Continue pressing SET> until ....
DIAG AIO
AIN CALIBRATED: NO
<SET
CAL
DIAG AIO
CALIBRATING A/D ZERO
DIAG AIO
DIAG AIO
AIN CALIBRATED: YES
<SET
CAL
From the
ENTR
CALIBRATING A/D SPAN
Operating Instructions
EXIT
EXIT
EXIT
EXIT
This
153