APPENDIX A-3: Warnings and Test Measurements

1
Name
WSYSRES
WDATAINIT
WCONFIGINIT
9
WNOXALARM1
9
WNOXALARM2
9
WNOALARM1
9
WNOALARM2
9
WNO2ALARM1
9
WNO2ALARM2
5+9
WO2ALARM1
5+9
WO2ALARM2
8+9
WCO2ALARM1
8+9
WCO2ALARM2
WSAMPFLOW
WOZONEFLOW
WOZONEGEN
WRCELLPRESS
WBOXTEMP
WRCELLTEMP
4
WMANIFOLDTEMP
8
WCO2CELLTEMP
5
WO2CELLTEMP
WIZSTEMP
WCONVTEMP
WPMTTEMP
WAUTOZERO
11
WPREREACT
07270D DCN7141
Table A-2:
Warning Messages
Message Text
Warnings
SYSTEM RESET
DATA INITIALIZED
CONFIG INITIALIZED
NOX ALARM 1 WARN
NOX ALARM 2 WARN
NO ALARM 1 WARN
NO ALARM 2 WARN
NO2 ALARM 1 WARN
NO2 ALARM 2 WARN
O2 ALARM 1 WARN
O2 ALARM 2 WARN
CO2 ALARM 1 WARN
CO2 ALARM 2 WARN
SAMPLE FLOW WARN
OZONE FLOW WARNING
OZONE GEN OFF
RCELL PRESS WARN
BOX TEMP WARNING
RCELL TEMP WARNING
MANIFOLD TEMP WARN
CO2 CELL TEMP WARN
O2 CELL TEMP WARN
IZS TEMP WARNING
CONV TEMP WARNING
PMT TEMP WARNING
AZERO WRN XXX.X MV
11
PRACT WRN XXX.X MV
APPENDIX A-3: Warnings and Test Measurements
Description
Instrument was power-cycled or the CPU
was reset.
Data storage was erased.
Configuration storage was reset to factory
configuration or erased.
NO
concentration alarm limit #1
X
exceeded
NO
concentration alarm limit #2
X
exceeded
NO concentration alarm limit #1 exceeded
NO concentration alarm limit #2 exceeded
NO
concentration alarm limit #1
2
exceeded
NO
concentration alarm limit #2
2
exceeded
O
concentration alarm limit #1 exceeded
2
O
concentration alarm limit #2 exceeded
2
CO
concentration alarm limit #1
2
exceeded
CO
concentration alarm limit #2
2
exceeded
Sample flow outside of warning limits.
Ozone flow outside of warning limits.
Ozone generator is off. This is the only
warning message that automatically
clears itself. It clears itself when the ozone
generator is turned on.
Reaction cell pressure outside of warning
limits.
Chassis temperature outside of warning
limits.
Reaction cell temperature outside of
warning limits.
Bypass or dilution manifold temperature
outside of warning limits.
CO
sensor cell temperature outside of
2
warning limits.
O
sensor cell temperature outside of
2
warning limits.
IZS temperature outside of warning limits.
Converter temperature outside of warning
limits.
PMT temperature outside of warning
limits.
Auto-zero reading above limit. Value
shown in message indicates auto-zero
A-9