Table 4-25: Current Loop Output Calibration With Resistor - Teledyne T200H User Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Operating Instructions
FULL SCALE
0%
100%
From the
AIO CONFIGURATION SUBMENU
(in the SETUP DIAG menu)
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
DIAG AIO
AOUTS CALIBRATED: NO
SET>
CAL
Continue pressing SET> until you reach the
output to be configured
DIAG AIO
DATA_OUT_2: CURR, NXCNC1, NOCAL
<SET SET>
EDIT
DIAG AIO
DATA_OUT_2 RANGE: CURR
SET>
EDIT
Continue pressing SET> until ...
DIAG AIO
DATA_OUT_2 CALIBRATED:NO
<SET SET>
CAL
152

Table 4-25: Current Loop Output Calibration with Resistor

VOLTAGE FOR 2-20 MA
(measured across 250Ω resistor)
0.5 V
5.0 V
To adjust the zero and span values of the current outputs, press:
ENTR
EXIT
EXIT
EXIT
EXIT
EXIT
Teledyne API - Model T200H/T200M Operation Manual
VOLTAGE FOR 4-20 MA
(measured across 250Ω resistor)
DIAG AIO
DATA_OUT_2 CURR-Z: 0 mV
U100
UP10
UP
DOWN DN10 D100 ENTR EXIT
EXAMPLE
DIAG AIO
DATA_OUT_2 CURR-Z:
U100 UP10 UP
DOWN DN10 D100
DIAG AIO
DATA_OUT_2 CURR-S: 5000 mV
U100 UP10 UP
DOWN DN10
D100
EXAMPLE
DIAG AIO
DATA_OUT_2 CURR-S: 4866 mV
U100 UP10 UP
DOWN DN10 D100
DIAG AIO
DATA_OUT_2 CALIBRATED: YES
<SET SET> CAL
1.0 V
5.0 V
Increase or decrease
the current output by
100, 10 or 1 counts.
The resulting change in
output voltage is
13 mV
displayed in the upper
line.
ENTR
EXIT
Continue adjustments
until the correct current
is measured with the
current meter.
ENTR EXIT
ENTR EXIT
EXIT
07270D DCN7141