CALIBRATION
CHANGING CALIBRATION FACTORS FOR MANUAL ASPIRATION MODE
9.
Press
10. Repeat steps 6 through 10 for each value you want to change.
11. Press one of the following:
SYSTEM RUN
RETURN
12. Set number of aspirations back to 01.
1.9
CHANGING CALIBRATION FACTORS FOR MANUAL ASPIRATION MODE
1.
Go to the Analyzer screen.
2.
Press
MAIN MENU
3.
Press
ANALYZER FUNCTIONS
4.
Press
CALIBRATION
5.
Press
MANUAL CALIBRATION FACTORS
Analyzer screen.
6.
Press the button corresponding to the parameter you want to change. Example: Press the
button next to WBC.
7.
Go to the Numeric Keypad.
8.
Use the keypad to provide the desired calibration factor. Example: Press
9
9.
Press
10. Repeat steps 6 through 9 for each value you want to change.
11. Press one of the following:
SYSTEM RUN
RETURN
1-10
E
N
T
E
R
.
.
.
.
6
6
.
E
N
T
E
R
.
To cycle samples. The next sample you cycle automatically
updates the calibration factors on the Workstation.
To display the CALIBRATION MENU screen.
To cycle samples. The next sample you cycle automatically
updates the calibration factors on the Workstation.
To display the CALIBRATION MENU screen.
. The existing calibration factors appear on the
.
0
PN 773023AC