TROUBLESHOOTING
DILUTER FUNCTIONS
3.
Press
4.
Press
Refilling the Reagent Reservoirs
1.
Go to the Numeric Keypad.
2.
Press
displays REFILLING ... (whatever reagent is low) while the instrument performs this
function. When the function completes processing, the Numeric Keypad displays
FUNCTION = 31. Press
Note: If the reservoirs are already primed, RESERVOIRS FULL PRESS ALARM RESET
appears on the Numeric Keypad. Press
Numeric Keypad.
3.
Press
Dispensing Retic Stain
1.
Go to the Numeric Keypad.
2.
Press
Numeric Keypad displays RETIC STAIN while the instrument performs this function.
When the function completes processing, the Numeric Keypad displays
FUNCTION = 25. Press
3.
Press
5-24
E
N
T
E
R
to perform this function again.
STOP
twice to exit this function. The Numeric Keypad displays READY.
E
N
T
E
F
3
1
R
to prime the reagent reservoirs. The Numeric Keypad
E
N
T
E
R
to perform this function again.
STOP
to exit this function. The Numeric Keypad displays READY.
E
N
T
E
F
2
5
R
to dispense Retic stain into the mixing chamber. The
E
N
T
E
R
to perform this function again.
STOP
to exit this function. The Numeric Keypad displays READY.
ALARM
RESET
and FUNCTION = 31 displays on the
PN 773023AC