Beckman Coulter LH 780 Instructions Manual page 186

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Table 6.5 System Messages - D (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Diluter Echo Error
NOT READY
N/A
Diluter Error
NOT READY
DILUTER ERROR
Diluter Left Power Sup.
(current status)
N/A
Bad
Diluter RAM Fault
SYSTEM FAULT
DILUTER RAM
FAULT
Diluter Right Power
(current status)
N/A
Sup. Bad
Diluter TxRx Error
NOT READY
N/A
Door Open Run Stopped (current status)
DOOR OPEN
RUN STOPPED
6-22
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
No
Instrument Log
Data displayed if available
(depends upon when error
Diluter communication
occurred).
failure. Reset Analytical
Station.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Diluter failure. Reset
occurred).
Analytical Station.
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
Diluter left power supply
Alarm" message printed in
defective. Reset Analytical
footer of sample report
Station.
(depends upon when error
occurred).
No
Instrument Log
Data not transmitted.
Diluter memory failure.
Reset Analytical Station.
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
Diluter right power supply
Alarm" message printed in
defective. Reset Analytical
footer of sample report
Station.
(depends upon when error
occurred).
No
Instrument Log
N/A
Communication failure
during transmission to
diluter. Reset Analytical
Station.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Diluter front door open
occurred).
during automatic
sampling mode operation.
Close front door and
continue operation.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
Diluter echo not understood by
Analyzer.
Stopped - inoperable.
Generic fatal Diluter error.
Stopped - inoperable.
Diluter left power supply
defective.
Stopped - inoperable.
Internal memory error.
Stopped - inoperable.
Diluter right power supply
defective.
Stopped - inoperable.
Communications error during
transmission to the Diluter.
Run stopped.
Diluter front door open during
Automatic mode operation.
6-22
Corrective Action
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
Close Diluter front door and continue operation.
PN 773023AC