TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Table 6.4 System Messages - C (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Compressor Too Hot
READY
COMPRESSOR
TOO HOT
CRC Error
NOT READY
COMMON ROM
AREA CRC ERROR
CRC Error
NOT READY
OVERLAY AREA
CRC ERROR
6-16
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
Pneumatics/Power Supply
Alarm" message printed in
overheated or defective.
footer of sample report
Verify air filter is clean.
(depends upon when error
occurred).
Yes
Instrument Log
N/A
Faulty diluter memory.
Reset Analytical Station.
Yes
Instrument Log
N/A
Faulty diluter memory.
Reset Analytical Station.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
Overheated pneumatics and/or
Pneumatic Power Supply
If this error occurs it will
defective (compressor shuts
not halt the action in
OFF when compressor
process, but it will prevent
temperature >80°C).
initiation of a new sample,
unless the error condition
is corrected.
Stopped - inoperable.
Incorrect or faulty Diluter
PROM chip.
Stopped - inoperable.
Incorrect or faulty Diluter
PROM chip.
6-16
Corrective Action
1. Turn the Power Supply MAIN POWER switch OFF
for 5 minutes.
2. Verify air filter is clean.
3. Turn the Power Supply MAIN POWER switch ON.
4. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
PN 773023AC