Table 6.6 System Messages - E through G (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Flow Cell Clog
SYSTEM FAULT
FLOW CELL CLOG
and
NOT READY
Function Not Allowed
(current status)
N/A
PN 773023AC
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
Yes
Instrument Log
Data for VCS-affected
parameters incomplete; colons
Instrument detected clog
(:::::) appear in VCS-affected
in flow cell and attempted
parameter fields if clog is
three times to clear clog
cleared after autoclearing. If
without success. Perform
clog is not cleared, data will not
F44 function.
be available.
No
Instrument Log
N/A
AMC Reported Error -
Function Not Allowed.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
Analytical Station detected clog
in flow cell and attempted three
times to clear clog without
success.
Paused until correct
F58 or F59 selected and
function entered.
Analytical Station does not
allow CD4 or CD8
(COMMUNICATION INTERFACE
card switches).
TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Corrective Action
1. Reset the Analytical Station.
2. Perform F44 FLOWCELL UNCLOG 1 function.
3. If message recurs, repeat F44 FLOWCELL UNCLOG
1 function.
4. If message recurs, perform F45 FLOWCELL
UNCLOG 2 function, up to three times.
CAUTION
Risk of instrument damage. Using F46
can damage the flow cell because it places greater
stress on the flow cell than F44 or F45. Only use F46
if F44 and F45 fail to clear the clog, and never do F46
more than once unless specifically instructed to do
so by your Beckman Coulter Representative.
5. If message recurs, perform F46 FLOWCELL
UNCLOG 3 function once.
6. Allow Analytical Station to sit idle for 30 minutes.
Then perform F44 FLOWCELL UNCLOG 1 function
before analyzing samples.
7. If message persists, call your Beckman Coulter
Representative.
Enter correct function for Analytical Station's mode of
operation.
6-25