H. Replace the aperture bath and reconnect it.
I.
N. Run a control in CBC test mode.
Procedure
DRAIN
A Press
POWER OFF
B Press
C Turn Off the Main Breaker on the Power Supply
WARNING
Risk of personal injury. A shock hazard exists if the power cord is connected. Unplug the
primary power cord before performing these procedures.
D Disconnect the Primary Power Cord
PN 773023AC
START UP
.
Until the Aperture Baths Drain Completely into the Waste Chamber
REPLACEMENT PROCEDURES
REPLACING AN APERTURE BATH
3
3-3