ADJUSTMENT PROCEDURES
ADJUSTING HGB LAMP VOLTAGE
Note: The LH compressor should not be manually turned off until all slide processing
has been completed at the SlideMaker.
4.4
ADJUSTING HGB LAMP VOLTAGE
1.
Verify that the baths contain diluent, not cleaning agent.
2.
Go to the Analyzer screen.
3.
Press
MAIN MENU
4.
Press
ANALYZER FUNCTIONS
5.
Press
HGB LAMP ADJUST
Analyzer. The message Adjusting HGB appears on the Keypad display. (If the voltage
appears out of range, call your Beckman Coulter Representative.
6.
Press
SYSTEM RUN
7.
Cycle three (3) samples of diluent in CBC Manual aspiration mode. Ignore all results,
including Hgb "....." incomplete computation.
4.5
SETTING UP BAR-CODE CONFIGURATION
Note: Use this procedure to indicate the number of digits only when you use Interleaved
2-of-5 bar-code labels for sample tubes.
1.
Go to the Analyzer screen.
2.
Press
MAIN MENU
3.
Press
SYSTEM CONFIGURATION
4.
Press
NUMBER OF BAR-CODE DIGITS (2 OF 5) XX
bar-code digits.
5.
Go to the Numeric Keypad. It displays VALUE? XXXX
6.
Use the Keypad to specify the number of digits (from 03 through 11).
7.
Press
8.
Go to the Analyzer screen.
9.
Confirm the number you specified appears on the Analyzer screen.
10. Press one of the following:
SYSTEM RUN
RETURN
4-2
.
.
. The message Performing HGB Lamp Adjust appears on the
.
.
.
E
N
T
E
R
after specifying the number of digits.
To cycle samples.
To display the Analyzer MAIN MENU screen.
, where
XX
represents the current number of
PN 773023AC