Table 6.9 System Messages - L
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Label Not Read Run
(current status)
LABEL NOT READ
Stopped
RUN STOPPED
Laser Voltage Lower
(current status)
N/A
Failed
Laser Voltage Too High
NOT READY
N/A
Laser Voltage Too Low
NOT READY
N/A
PN 773023AC
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
Yes
Instrument Log
Data not available.
Unable to read cassette
label. Check label and
scanner. Rerun sample.
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Laser voltage below limit
Alarm" message printed in
(<Tube ID>, <Cass/Pos>).
footer of sample report
(depends upon when error
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Laser voltage too high.
Alarm" message printed in
footer of sample report
(depends upon when error
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Laser voltage too low.
Alarm" message printed in
footer of sample report
(depends upon when error
occurred).
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Run stopped
Unable to read cassette label.
If AutoStop/Flow Cell
Voltage is 1181 to 1230 Vdc.
switch is OFF, run
continues.
If AutoStop/Flow Cell
switch is ON, run is
stopped at end of cycle.
Voltage ≥1520 Vdc.
Idle - ready for request.
Voltage ≤ 1180 Vdc.
Idle - ready for request.
TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Corrective Action
1. Clean the bar-code reader.
2. Check the cassette label and clean or replace the
label if necessary.
3. Continue operation. If message persists, use a
different cassette.
4. If message still persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. Continue normal operation and watch for additional
messages.
3. If additional messages appear, follow the corrective
actions associated with the messages.
4. If this message persists:
a. Change the test mode to a non-Diff test.
b. Continue processing non-Diff samples.
c. Call your Beckman Coulter Representative.
1. Reset the Analytical Station.
2. Continue normal operation and watch for additional
messages.
3. If additional messages appear, follow the corrective
actions associated with the messages.
4. If this message persists:
a. Change the test mode to a non-Diff test.
b. Continue processing non-Diff samples.
c. Call your Beckman Coulter Representative.
1. Reset the Analytical Station.
2. Continue normal operation and watch for additional
messages.
3. If additional messages appear, follow the corrective
actions associated with the messages.
4. If this message persists:
a. Change the test mode to a non-Diff test.
b. Continue processing non-Diff samples.
c. Call your Beckman Coulter Representative.
6-29
6