Table 6.5 System Messages - D (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
DPR Diluter Packet
SYSTEM FAULT
N/A
Error
and
NOT READY
Dump Valve Stuck
SYSTEM FAULT
DUMP VALVE
Closed
STUCK CLOSED
PN 773023AC
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
Yes
Instrument Log
Data displayed if available
(depends upon when error
Invalid dual port RAM
occurred).
packet (<Tube ID>,
<Cass/Pos>). Reset
Analytical Station and
rerun sample.
Yes
Instrument Log
Data not transmitted.
Overflow valve
malfunction (remained
closed) detected while
turning compressor ON.
Reset Analytical Station.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Idle - ready for request or
DPR malfunction.
cycle continues.
Stopped - inoperable.
Power Supply valve
malfunction.
TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Corrective Action
1. Continue normal operation and watch for additional
messages.
2. If additional messages appear, follow the corrective
actions associated with those messages.
3. If this message persists:
a. Change the test mode to CBC Only.
b. Continue processing CBC only samples.
c. Call your Beckman Coulter Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
6-23