REPLACEMENT PROCEDURES
REPLACING TUBING
45.75 cm (18")
8.
Tighten the caps.
9.
Insert the plastic collars that secure the pickup tubes.
10. Record the new containers' information in Quality Assurance Set Up, in your laboratory
logbook and on the reagent package.
11. Ensure the pneumatics are on.
12. Press
3.13 REPLACING TUBING
WARNING
Possible biohazardous condition. Tubing can come in contact with biohazard materials. Skin
contact with these materials can be harmful to your health. Handle and dispose of tubing in accordance with
your laboratory's procedures.
CAUTION
Risk of damage to tubing. Using alcohol or any liquid to lubricate the tubing when you replace it
can damage the tubing. Do not lubricate the tubing when you replace it.
IMPORTANT
damaged, replace it with a new piece of tubing of the same type and length. If you need tubing, call your
Beckman Coulter Representative.
3-26
45.75 cm (18")
START
CONT
from the Numeric Keypad to continue cycling samples.
Incorrect results can occur if you change the length of tubing. If tubing tears or becomes
3.
Remove the plastic collar that secures
the pickup tube assembly.
4.
Unscrew the pickup tube assembly from
the old container.
5.
Lift the assembly straight up and out.
IMPORTANT
Incorrect results can occur if the
tubes become contaminated. Do not touch the
tubes or let them touch any laboratory surfaces. If
the tubes touch anything, rinse them with distilled
water and then wipe them with a lint-free tissue.
6.
Inspect the pickup tubes for damage,
and replace them if necessary.
7.
Carefully insert each end of the pickup
tube assembly straight into the new
containers.
PN 773023AC