TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Table 6.16 System Messages - W through Z
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Workstation
NOT READY
N/A
Communications
Timeout
Wrong Cal. Factors
NOT READY
N/A
Wrong Command
NOT READY
INVALID COMMAND
Interpretation
XX RECEIVED
Wrong Command
NOT READY
NOT READY FOR XX
Sequence
COMMAND AFTER
YY
Wrong Echo Xmit
NOT READY
N/A
6-48
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
No
No
Instrument Log
Data not transmitted.
Incorrect calibration
factors or bath voltages.
Re-enter calibration
factors and/or bath
voltages.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Diluter received an invalid
occurred).
command from the AMC.
Reset Analytical Station.
Yes
N/A
Data displayed if available
(depends upon when error
occurred).
No
Instrument Log
Data not transmitted.
AMC-DIFF communication
failure: wrong echo. Reset
Analytical Station.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped
Communication timeout.
Communication between the
Analytical Station and the
Workstation is lost.
Stopped - inoperable.
Incorrect calibration factors or
bath voltages.
Stopped - inoperable.
Invalid command received by
Diluter.
Stopped - inoperable.
Communication error.
Stopped - inoperable.
Wrong echo received from
Diluter while transmitting data.
6-48
Corrective Action
1. Reset the Analytical Station.
2. Continue normal operation. If message recurs, call
your Beckman Coulter Representative.
1. Reset the Analytical Station.
2. Re-enter the calibration factors and/or aperture
voltages.
3. Continue normal operation. If the message recurs,
call your Beckman Coulter Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
PN 773023AC