C Clean Bsv, - Beckman Coulter LH 780 Instructions Manual

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

C Clean BSV
PN 773023AC
CLEANING BLOOD SAMPLING VALVE (REMOVE AND CLEAN INSIDE)
TOP VIEW
CLEANING PROCEDURES
1.
Clean the guide post and mounting
post with distilled water and dry
them with a lint-free tissue.
CAUTION
Risk of damage to ceramic
sections can occur if they come in contact
with an abrasive material. Do not scratch or
use abrasive materials on the inside surfaces.
Never try to remove debris by inserting an
object in the valve¢s passageways.
2.
Hold a container or several lint-free
tissues beneath the BSV mounting
post to catch overflow.
3.
Clean the passageways, inside
surfaces, and center holes of the
three sections with a stream of
cleaning agent from a wash bottle.
4.
Place the spout of the wash bottle
over each opening. Force cleaning
agent into each passageway to free
and remove any debris or blood.
5.
If a stubborn plug remains lodged in
passageway, try a jet of filtered
compressed air to dislodge it.
6.
Use a lint-free tissue moistened with
distilled water to remove any debris
or dried blood that still remains on
the surface or in center holes.
7.
Rinse the passageways and inside
surfaces thoroughly with distilled
water.
2
2-19