7.
Press
8.
Remove the reagent line from the distilled water container and place it in the CLEANER
container.
9.
Press
10. Perform daily startup.
2.10 CLEANING BLOOD SAMPLING VALVE DRIP PLATE
WARNING
Possible biohazardous condition. The drip plate can contain residual biological materials and
must be handled with care. Avoid skin contact. Clean up spills immediately in accordance with your local
regulations and acceptable laboratory procedures. This may include, but is not limited to, wearing
protective eyewear, gloves, and suitable laboratory attire when operating or maintaining this analyzer.
PN 773023AC
STOP
twice to exit this function. The Numeric Keypad displays READY.
SHUT DOWN
.
CLEANING PROCEDURES
CLEANING BLOOD SAMPLING VALVE DRIP PLATE
1.
Prepare a bleach solution in a wash
bottle.
2.
Open the Diluter lower front door.
3.
Slide the drip plate out.
4.
Use the wash bottle to fill the drip
plate with the bleach solution. Be
careful not to overflow the drip
plate.
5.
Leave the bleach solution in the
drip plate for
6.
Use gauze held with a hemostat to
clean the entire surface.
7.
Use clean, dry gauze to soak up any
excess bleach solution.
8.
Use a wash bottle with distilled
water to rinse the drip plate.
9.
Use clean, dry gauze to soak up any
excess water.
10. Slide the drip plate back.
11. Close the Diluter lower front door.
2
.
2-23