TROUBLESHOOTING
WORKSTATION MESSAGES
Table 6.17 Workstation Messages - A through B (Continued)
Workstation
Log Message and Icon on Command Center
AutoStop has been triggered for control folder <lot#>.
Background CBC test values within limits
Background DIFF test values within limits
Background: <Background test/parameter> <Background>
exceeded <flag> limit
Body Fluid is not allowed in automatic sampling mode. Sample
will not be stored.
6-50
Applicable
Log
Sample Results
System Status
Probable Cause
Controls
N/A
Ready
A control Autostop condition has occurred.
Daily Checks
Ready
The background test passed for the test specified.
Daily Checks
Ready
The background test passed for the test specified.
Daily Checks
N/A
N/A
During a background test, one of the test parameters failed.
Test parameters are HGB blank, HGB read, WBC, RBC, HGB,
PLT, DIFF, RETIC, sheath pressure, DIFF pressure, DIFF LS
offset, RETIC LS offset, Laser voltage, RF300 Volts,
RF6.3 Volts, and ambient temperature.
Event
N/A
Ready
This message displays when a body fluid type is processed in
automatic sampling mode rather than manual mode.
System Status/Recovery
Corrective Action
1. Review the Control History Log to review specific error.
2. Perform corrective action per lab protocol.
N/A
N/A
1. Repeat the test.
2. If the message persists, call your Beckman Coulter Representative.
1. Rerun the sample in the manual aspiration mode.
6-50
PN 773023AC