TROUBLESHOOTING
DILUTER FUNCTIONS
Note: This function takes approximately 2 minutes.
When the function completes processing, the Numeric Keypad displays FUNCTION = 15.
Press
3.
Press
Priming Diluent
1.
Go to the Numeric Keypad.
2.
Press
PRIME DILUENT while the instrument performs this function.
Note: This function takes approximately 2 minutes.
When the function completes processing, the Numeric Keypad displays FUNCTION = 16.
Press
3.
Press
Priming Retic Clearing Solution
1.
Go to the Numeric Keypad.
2.
Press
instrument:
Dispenses Retic clearing solution to the Retic chamber
r
Drains the Retic chamber to the waste chamber.
r
The Numeric Keypad displays PRIME RETIC CLEAR while the instrument performs this
function.
Note: This function takes approximately 1 minute.
5-20
E
N
T
E
R
to perform this function again.
STOP
to exit this function. The Numeric Keypad displays READY.
E
N
T
E
F
1
6
R
to prime the diluent lines. The Numeric Keypad displays
E
N
T
E
R
to perform this function again.
STOP
to exit this function. The Numeric Keypad displays READY.
E
N
T
E
F
1
5
R
to prime Retic clearing solution reagent lines. The
PN 773023AC