3.14 REPLACING WASTE CONTAINER
WARNING
Possible biohazardous condition. The contents of the old waste container and its associated tubing
can include residual biological material and must be handled with care. Check the tubing connection and
container location periodically. Avoid skin contact and clean up spills immediately. Dispose of the contents of
the waste container in accordance with your local regulations and acceptable laboratory procedures.
PN 773023AC
REPLACEMENT PROCEDURES
REPLACING WASTE CONTAINER
1.
Carefully push the tubing straight onto
the fitting.
2.
Ensure it is securely in place.
Unscrew the pickup tube assembly from the
old container.
Lift the pickup tube assembly straight out of
the old container.
Place the pickup tube assembly straight into
the new container and screw it into position.
Label the new waste container and verify
that the old container is clearly labeled, then
discard old container according to your
laboratories standards for biohazardous
material.
3
3-27