TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Table 6.14 System Messages - T through V
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Two NAKs From Diluter
NOT READY
N/A
Two NAKs Sent From
NOT READY
N/A
AMC
6-46
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
No
Instrument Log
Data displayed if available
(depends upon when error
Analyzer-Diluter
occurred).
Communication failure:
two NAKs sent from
Diluter. Reset Analytical
Station.
No
Instrument Log
Data displayed if available
(depends upon when error
Analyzer-Diluter
occurred).
communication failure:
two NAKs sent from AMC.
Reset Analytical Station.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
Two NAKs detected while
receiving data from Diluter.
Stopped - inoperable.
Two NAKs detected while
transmitting data to Diluter.
6-46
Corrective Action
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
PN 773023AC