TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Table 6.10 System Messages - M through O (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Not Ready For AMC
NOT READY
NOT READY
Data
FOR AMC DATA
Operator Alert 09
NOT READY
N/A
6-34
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
Yes
Instrument Log
Data displayed if available
(depends upon when error
Diluter not ready to
occurred).
receive data from AMC.
Reset Analytical Station.
Yes
Instrument Log
Data not transmitted.
AMC Reported Error --
Operator Alert 09.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
Communication error.
Stopped - inoperable.
Miscommunication between the
Analyzer and the Diluter.
6-34
Corrective Action
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station and reposition the
cassette.
2. If message persists, call your Beckman Coulter
Representative.
PN 773023AC