TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Table 6.11 System Messages - P through Q (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Probe Wipe Truck Not
SYSTEM FAULT
PROBE WIPE
Home
TRUCK
and
NOT HOME
NOT READY
Probe Wipe Up Fail
SYSTEM FAULT
PROBE WIPE UP
FAIL
and
NOT READY
PS1 +5V Supply High
SYSTEM FAULT
N/A
PS1 +5V Supply Low
SYSTEM FAULT
N/A
PS1 +15V Supply High
SYSTEM FAULT
N/A
PS1 +15V Supply Low
SYSTEM FAULT
N/A
PS1 -15V Supply High
SYSTEM FAULT
N/A
6-36
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
Yes
Instrument Log
Data displayed.
Manual sampling probe
wipe obstructed (not
home) (<Tube ID>,
<Cass/Pos>). Remove
obstruction.
Yes
Instrument Log
Data displayed.
Manual sampling probe
wipe obstructed (not up)
(<Tube ID>, <Cass/Pos>).
Remove obstruction.
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 1 problem:
Alarm" message printed in
+5 Volts above limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 1 problem:
Alarm" message printed in
+5 Volts below limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 1 problem:
Alarm" message printed in
+15 Volts above limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 1 problem:
Alarm" message printed in
+15 Volts below limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 1 problem:
Alarm" message printed in
-15 Volts above limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
r Sensor is obstructed or
faulty
r Probe-wipe mechanism is
obstructed or faulty.
Stopped - inoperable.
Obstruction or faulty sensor.
Stopped - inoperable.
+5 V power to power supply 1
outside +5% limit.
Stopped - inoperable.
+5 V power to power supply 1
outside -5% limit.
Stopped - inoperable.
+15 V power to power supply 1
outside +5% limit.
Stopped - inoperable.
+15 V power to power supply 1
outside -5% limit.
Stopped - inoperable.
-15 V power to power supply 1
outside +5% limit.
6-36
Corrective Action
1. Remove obstruction.
2. Reset the Analytical Station.
3. If message persists, call your Beckman Coulter
Representative.
1. Remove obstruction.
2. Reset the Analytical Station.
3. If message persists, call your Beckman Coulter
Representative.
Call your Beckman Coulter Representative.
Call your Beckman Coulter Representative.
Call your Beckman Coulter Representative.
Call your Beckman Coulter Representative.
Call your Beckman Coulter Representative.
PN 773023AC