REPLACEMENT PROCEDURES
REPLACING NEEDLE
B Attach Safety Clip and Remove Needle Cartridge
WARNING
Skin puncture from the needle, which might contain biohazardous material, can occur. Handle
the needle cartridge with extreme care. Always use the safety clip to remove and install the needle cartridge.
This protects you from possible needle puncture
C Clean Needle Assembly Area
3-10
.
1.
Without squeezing the safety clip, fit
the right edge of the safety clip into
the groove on the right side of the
front support of the needle cartridge.
2.
Slide the safety clip to the left until its
left edge enters the groove on the left
side of the front support of the needle
cartridge. You should hear a click.
CAUTION
Risk of system damage. Pulling the
needle cartridge out too far can tear the
tubing. Pull the cartridge out slowly.
3.
Pull out the needle cartridge.
1.
If there is blood on the cover:
a.
Clean the cover with a cotton
swab moistened with a 50:50
solution of bleach and
distilled water.
b.
Rinse another cotton swab in
distilled water and wipe the
same area.
2.
Dip a cotton swab into a 50:50
solution of bleach and distilled
water to clean upper and lower
grooves where the needle assembly
fits.
3.
Rinse another cotton swab in
distilled water and wipe the same
area.
PN 773023AC