TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Table 6.11 System Messages - P through Q (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
PS2 -15V Supply High
SYSTEM FAULT
N/A
PS2 -15V Supply Low
SYSTEM FAULT
N/A
P/W Harness Not
(current status)
P/W HARNESS NOT
Connected
CONNECTED
6-38
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 2 problem:
Alarm" message printed in
-15 Volts above limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 2 problem:
Alarm" message printed in
--15 Volts below limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
Yes
Instrument Log
Processed normally.
The probe wipe harness is
not connected. Reset
Analytical Station.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
-15 V power to power supply 2
outside +5% limit.
Stopped - inoperable.
-15 V power to power supply 2
outside -5% limit.
Beeps- operation
Harness not connected.
continues.
If this error occurs it will
not halt the action in
process, but it will prevent
initiation of a new sample,
unless the error condition
is corrected.
6-38
Corrective Action
Call your Beckman Coulter Representative.
Call your Beckman Coulter Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
PN 773023AC