CLEANING PROCEDURES
CLEANING MANUAL ASPIRATION PROBE WIPE
2.13 CLEANING MANUAL ASPIRATION PROBE WIPE
WARNING
Possible biohazardous condition. The manual aspiration probe wipe can contain residual
biological materials and must be handled with care. Avoid skin contact. Clean up spills immediately in
accordance with your local regulations and acceptable laboratory procedures. This may include, but is not
limited to, wearing protective eyewear, gloves, and suitable laboratory attire when operating or maintaining
this analyzer.
2.14 BLEACHING RBC INTERNAL ELECTRODES
Summary
CAUTION
Risk of system damage. Do not mix diluent with bleach. A precipitate forms that can clog the
apertures.
A.
Prepare 50 mL of bleach solution in a wash bottle.
B.
Put the bleach solution in the WBC bath.
C. Put the bleach solution in the RBC bath.
D. Perform Extended Prime.
E.
Disconnect the sweep flow lines and attach tubing to draw up bleach.
2-26
1.
Open the Diluter lower front door.
2.
Open the black activator.
3.
Use distilled water to rinse the
probe wipe.
4.
Use a lint-free tissue to remove
any dried blood and dry the probe
wipe.
5.
Close the black activator.
6.
Close the Diluter lower front door.
PN 773023AC