Sort Mode Library

The Sort Mode Library is used to manage the sort modes, and set the default sort mode for sorting.
To customize a sort mode, you have to add a sort mode first. Refer to
Select
Advanced
Figure 2.12 Sort Mode Library
NOTE
The Sort Mode Library can be edited only when an experiment is closed.
NOTE
Select Default for Tube or Default for Plate to set the most frequently used Sort Mode as the default
for the tube sorting or plate sorting.

Sort Mode

Sort Mode is a combination setting of Sort Precision Level and Drop Envelope. Four sort modes
including
Single
changed.
The target population may be sorted using four different modes depending on the output desired.
Each mode is defined by the location of the particle within the drop and the location of
contaminants in either the primary drop (the central drop of interest for the sort logic) or an
adjacent drop (a single drop on either side of the primary drop).
Single Mode
sort.
Purity Mode
C37808AC
>
to access the Sort Mode Library.

Sort Mode Library

,
,
and
Purity
Purity 1-2
Enrich
: Used when having only one event per drop is the most important aspect of the
: Used when the purity of the sort is the most important.
Using the CytExpert SRT Software
Figure
are available by default and cannot be deleted or
2
Sort Mode Library
2.12.
2-47