Beckman Coulter LH 780 Instructions Manual page 173

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

Table 6.2 System Messages - A (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
AMC Requested Stop
(current status)
AMC REQUESTED
STOP
Analyzer +48V Bad
READY
ANALYZER +48V
BAD
Analyzer Error
SYSTEM FAULT
ANALYZER ERROR
and
NOT READY
Analyzer Fans Failed
(current status)
N/A
Aperture Voltage Lower
(current status)
N/A
Failed
Aperture Voltage Upper
(current status)
N/A
Failed
PN 773023AC
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
No
Instrument Log
Data displayed if available
(depends upon when error
Analyzer stopped. Reset
occurred).
Analytical Station. If
message persists,
perform flow cell unclog
procedure.
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
48 Volt power outside
Alarm" message printed in
±10% limit.
footer of sample report
(depends upon when error
occurred).
Yes
Instrument Log
Data displayed if available
(depends upon when error
Communications error.
occurred).
Reset Analytical Station.
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
Analyzer fans power
Alarm" message printed in
supply defective. Reset
footer of sample report
Analytical Station.
(depends upon when error
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Aperture voltage reading
Alarm" message printed in
below limits.
footer of sample report
(depends upon when error
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Aperture voltage reading
Alarm" message printed in
above Limits.
footer of sample report
(depends upon when error
occurred).
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Paused until error
Analyzer needs to stop after an
resolved.
abnormal condition occurred
(for example, clearing flow cell
clogs).
Stopped - inoperable.
+48 V power to Analyzer
outside ±10% limit.
If this error occurs it will
not halt the action in
process, but it will prevent
initiation of a new sample,
unless the error condition
is corrected.
Stopped - inoperable.
Communication error.
Stopped - inoperable.
Analyzer fans power supply
defective.
If AutoStop/Flow Cell
Voltage reading <277 V
switch is OFF, run
(measured at the aperture) or
continues.
<4.332 V (measured at the edge
connector of the DIFF
If AutoStop/Flow Cell
PROCESSOR card).
switch is ON, run is
stopped at end of cycle.
If AutoStop/Flow Cell
Voltage reading >322 V
switch is OFF, run
(measured at the aperture) or
continues.
>5.045 V (measured at the edge
connector of the DIFF
If AutoStop/Flow Cell
PROCESSOR card).
switch is ON, run is
stopped at end of cycle.
TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Corrective Action
1. Reset the Analytical Station.
2. If message persists, perform the flow cell unclog
function (for example, F44, F45 or F46).
1. Reset the Analytical Station.
2. If message persists:
a. Turn Power Supply MAIN POWER switch OFF
for 5 minutes.
b. Verify air filter is clean.
c. Turn Power Supply MAIN POWER switch ON.
3. If message still persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
6-9