Table 6.13 System Messages - S (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Stripper Plate Not Out
SYSTEM FAULT
STRIPPER PLATE
NOT OUT
and
MANUAL MODE
STX Rec. Error
NOT READY
N/A
Sweep Flow Tank Not
SYSTEM FAULT
SWEEP FLOW TANK
Full
and
NOT FULL
NOT READY
Sweep Flow Tank Still
SYSTEM FAULT
SWEEP FLOW TANK
Full
and
STILL FULL
NOT READY
PN 773023AC
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
Yes
Instrument Log
Data displayed if available
(depends upon when error
Stripper plate obstructed
occurred).
(not out) or faulty sensor.
Remove obstruction and
restart Analytical Station.
No
Instrument Log
Data not transmitted.
Incorrect calibration
factor transmission.
Retransmit calibration
factors from GEN•S
Workstation.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Diluent level low or faulty
occurred).
tank sensor. Check diluent
and restart Analytical
station.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Diluent level low or faulty
occurred).
tank sensor.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Automatic mode disabled;
Obstruction or faulty sensor.
Manual mode operable.
Stopped - inoperable.
The first character not STX
while receiving K-CAL factors.
Stopped Inoperable
Diluent level low or faulty float
and/or sensor,
Stopped Inoperable
Faulty float and/or sensor.
Sensor valves inoperable or
defective.
TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Corrective Action
1. Ensure the stripper plate is unobstructed.
2. Reset the Analytical Station.
3. Continue normal operation. If the message recurs,
call your Beckman Coulter Representative. You can
run samples in the Manual mode.
1. Reset the Analytical Station.
2. Retransmit the calibration factors to the Analyzer.
3. Continue normal operation. If message recurs, call
your Beckman Coulter Representative.
1. Check diluent container for sufficient volume and
replace if necessary.
2. Reset the Analytical Station.
3. If the volume of diluent is adeqaute or message
recurs, check for leaks and kinks in the tubing
around the reservoir.
4. Continue operation. If message persists, call your
Beckman Coulter Representative.
1. Reset the Analytical Station.
2. If message persisits,
a. Use F50 to check operation of the Sweep Flow
Reservoir
b. check for kinks in the tubing around the
reservoir.
3. Continue operation. If message persists, call your
Beckman Coulter Representative.
6-45