Beckman Coulter LH 780 Instructions Manual page 203

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

Table 6.12 System Messages - R
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Red/White Reset Error
NOT READY
N/A
Retic Accumulation Not
NOT READY
N/A
Finished
Retic Laser Offset Lower
READY
N/A
Failed
Retic Laser Offset Upper
READY
N/A
Failed
Retic Pressure Lower
(current status)
N/A
Failed
Retic Pressure Upper
(current status)
N/A
Failed
Retic Xmit Failure
(current status)
N/A
PN 773023AC
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
Yes
Instrument Log
Data not transmitted.
Red/white processor
counter(s) did not
reinitialize. Reset
Analytical Station.
No
Instrument Log
Data not transmitted.
Incomplete Retic
collection. Reset
Analytical Station.
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Retic laser offset too low.
Alarm" message printed in
footer of sample report
(depends upon when error
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Retic laser offset too high.
Alarm" message printed in
footer of sample report
(depends upon when error
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Retic sample pressure
Alarm" message printed in
below limit (<Tube ID>,
footer of sample report
<Cass/Pos>).
(depends upon when error
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Retic sample pressure
Alarm" message printed in
above limit (<Tube ID>,
footer of sample report
<Cass/Pos>).
(depends upon when error
occurred).
No
Instrument Log
Data not transmitted.
AMC Reported Error -
Retic Transmit Failure.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
Red and white counter did not
reset to zero value.
Stopped - inoperable.
Error during retic data
accumulation.
Idle - ready for request.
Analytical Station not adjusted
properly.
Idle - ready for request.
Analytical Station not adjusted
properly.
If AutoStop/Flow Cell
Pressure reading <0.3 psi.
switch is OFF, run
continues.
If AutoStop/Flow Cell
switch is ON, run is
stopped at end of cycle.
If AutoStop/Flow Cell
Pressure reading >2.2 psi.
switch is OFF, run
continues.
If AutoStop/Flow Cell
switch is ON, run is
stopped at end of cycle.
Stopped - inoperable.
Analytical
Station-to-Workstation
transmission failure.
TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Corrective Action
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
If message persists, call your Beckman Coulter
Representative.
If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
6-39