TROUBLESHOOTING
WORKSTATION MESSAGES
Table 6.25 Workstation Messages - W through Z
Workstation
Log Message and Icon on Command Center
Workload has been automatically reset
Workload has been manually archived
Workload has been manually printed
Wrong <mode> Type, Latex sample will be not be stored.
Wrong <mode> Type, Sample will be stored as a Patient
Sample.
Wrong Repro Type, latex Sample will not be stored.
XB all batches were deleted
XB current batch was deleted
6-66
Applicable
Sample
Log
Results
System Status
Probable Cause
Event
N/A
Ready
The workload recording is automatically reset (restart at
zero) after 365 days.
Event
N/A
Ready
Operator initiated an archiving of the workload
recording.
Event
N/A
Ready
Operator initiated a printing of the workload recording.
Repro or
Discarded
Ready
The run received is of Latex type (i.e., F55, F56, F57, ...)
Carryover
and cannot be received in the designated modes. Mode
types include Reproducibility, and Carryover.
Calibration or
Stored as
Ready
When running a calibration test the operator changed
Carryover
patient
the aspiration mode (Automatic vs. Manual).
sample
Repro
N/A
Ready
The run received is of Latex type (ie., F55, F56, F57 ...)
and cannot be received in the Reproducibility mode.
XB
N/A
Ready
The operator cleared all XB batches.
XB
N/A
Ready
The operator deleted the current XB batch.
System Status/Recovery
Corrective Action
Status only: no action necessary.
Status only: no action necessary.
Status only: no action necessary.
Rerun the sample in the correct aspiration mode. (Automatic or Manual).
Rerun the sample in the correct aspiration mode. (Automatic or Manual).
Verify Workstation mode of operation is consistent with sample material.
Status only; no action necessary.
Status only; no action necessary.
6-66
PN 773023AC