CLEANING PROCEDURES
ADVANCED BAR-CODE READER
6.
After the sample ID appears on the Analyzer. Press
STOP
7.
Ensure the sample ID is correctly displayed at the Analyzer screen.
Cleaning the Scan Window (Welch Allyn
Note: The following information is reprinted from the Welch Allyn IT 3800/3900 Handheld
Linear Imager User's Guide by permission of Welch Allyn, Inc.
CAUTION
Risk of damage to the handheld scanner. Do not submerge the scanner in water. The scanner's
housing is not water tight. Do not use abrasive wipers or tissues on the scan window: abrasive wipers may
scratch the window. Never use solvents (alcohol or acetone) on the housing or window: solvents can
damage the finish or the window.
Scanning performance can degrade if the scan window is not clean. If the scan window is
visibly dirty, or if the scanner is not scanning well, clean the scan window with a soft cloth or
tissue dampened with water (or a mild detergent-water solution). If a detergent solution is
used, rinse with a clean tissue dampened with water only.
The scanner housing can also be cleaned the same way.
2.17 ADVANCED BAR-CODE READER
The Advanced Bar-Code Reader is a vision based bar-code reader that uses two Imaging
Bar-Code Decoders. One decoder reads the cassette label, the other decoder reads the sample
label.
The instrument activates the bar-code reader while piercing a sample tube in Automatic
aspiration mode.
IMPORTANT
labels could prevent the instrument from reading the bar-code labels. Ensure the bar-code labels are
undamaged. Ensure the bar-code labels conform to the specifications provided in the online help.
The instrument reads both the tube and cassette labels. If the instrument cannot read the tube
label, the instrument only reports the cassette label.
2-34
to reject the sample ID.
Risk of misidentification. Use of poor quality, dirty, improperly placed or damaged bar-code
E
N
T
E
R
to accept the sample ID or
®
IT 3800)