REPLACEMENT PROCEDURES
REPLACING PINCH VALVES
E Remove the Pinch Valve Tubing and Pinch Valve
3-22
1.
Disconnect the tubing that goes through
the pinch valve at the fitting closest to the
pinch valve.
2.
Attach a red shipping clip to hold open the
valve.
3.
Pull the tubing through the pinch valve.
4.
Remove the red shipping clip.
5.
Using the pinch valve wrench, turn the
pinch valve about 1/4-turn until it snaps
out.
PN 773023AC