CLEANING PROCEDURES
CLEANING VACUUM ISOLATOR CHAMBER (VIC)
B Bleach VIC
1/2 to 3/4
full
C Return VIC to Working Order
1.
Reconnect the original U shaped tubing to the VIC.
2.
Open the hemostat to unpinch the tubing at bottom of the VIC.
3.
Go to the Numeric Keypad.
4.
Press
5.
Use F05 to activate solenoid 16 to drain the VIC and waste chamber.
D Shut Down Your System
E Start Up Your System
2-38
P O W E R O N
.
1.
Prepare a bleach solution.
2.
Connect the bleach bottle tubing to the
VIC fitting where the U-shaped tubing
was disconnected.
CAUTION
Risk of system damage. Operating the
instrument when VIC is filled with bleach can
cause VIC to overflow. Do not operate
instrument when the VIC is filled with bleach
solution.
3.
Fill the VIC 1/2 to 3/4 full with bleach
solution.
4.
Disconnect the bleach bottle.
5.
Leave the bleach solution in the VIC for
.
PN 773023AC