Beckman Coulter LH 780 Instructions Manual page 201

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

Table 6.11 System Messages - P through Q (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
PS1 -15V Supply Low
SYSTEM FAULT
N/A
PS2 +5V Supply High
SYSTEM FAULT
N/A
PS2 +5V Supply Low
SYSTEM FAULT
N/A
PS2 +12V Supply High
SYSTEM FAULT
N/A
PS2 +12V Supply Low
SYSTEM FAULT
N/A
PS2 +15V Supply High
SYSTEM FAULT
N/A
PS2 +15V Supply Low
SYSTEM FAULT
N/A
PN 773023AC
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 1 problem:
Alarm" message printed in
-15 Volts below limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 2 problem:
Alarm" message printed in
+5 Volts above limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 2 problem:
Alarm" message printed in
+5 Volts below limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 2 problem:
Alarm" message printed in
+12 Volts above limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 2 problem:
Alarm" message printed in
+12 Volts below limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 2 problem:
Alarm" message printed in
+15 Volts Above limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
No
Instrument Log
Data not transmitted, or data
transmitted and "System
Power Supply 2 problem:
Alarm" message printed in
+15 Volts below limit. Call
footer of sample report
your Beckman Coulter
(depends upon when error
Representative.
occurred).
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
-15 V power to power supply 1
outside -5% limit.
Stopped - inoperable.
+5 V power to power supply 2
outside +5% limit.
Stopped - inoperable.
+5 V power to power supply 2
outside -5% limit.
Stopped - inoperable.
+12 V power to power supply 2
outside +5% limit.
Stopped - inoperable.
+12 V power to power supply 2
outside -5% limit.
Stopped - inoperable.
+15 V power to power supply 2
outside +5% limit.
Stopped - inoperable.
+15 V power to power supply 2
outside -5% limit.
TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Corrective Action
Call your Beckman Coulter Representative.
Call your Beckman Coulter Representative.
Call your Beckman Coulter Representative.
Call your Beckman Coulter Representative.
Call your Beckman Coulter Representative.
Call your Beckman Coulter Representative.
Call your Beckman Coulter Representative.
6-37