Table 6.21 Workstation Messages - L through M
Workstation
Log Message and Icon on Command Center
Last Two XB batches are outside limits.
Yellow traffic light is displayed if XB AutoStop option is
disabled.
Red traffic light is displayed if XB AutoStop option is
enabled.
Last XB Batch is out of limits.
Yellow traffic light is displayed if XB AutoStop option is
disabled.
Red traffic light is displayed if XB AutoStop option is
enabled.
Last XB batch outside limit conditions has cleared
Last CBC XM batch is out of limits.
Yellow traffic light is displayed if XM AutoStop option is
disabled.
Red traffic light is displayed if XM AutoStop option is
enabled.
Last CBC XM batch outside limits condition has cleared.
Last Diff XM batch is out of limits.
Yellow traffic light is displayed if XM AutoStop option is
disabled.
Red traffic light is displayed if XM AutoStop option is
enabled.
PN 773023AC
Applicable
Sample
Log
Results
System Status
Probable Cause
XB
N/A
Analytical Station
Last XB batch is outside limits.
stopped if XB
AutoStop option (2
batches out) is
enabled.
XB
N/A
Analytical Station
The last XB batch is outside limits.
stopped if XB
AutoStop option (1
batch out) is enabled.
XB
N/A
Ready
Last XB batch was out of limits and current XB batch is
within limits.
XM
N/A
Analytical Station
The last XM batch is outside limits.
stopped if XM
AutoStop option (1
batch out) is enabled.
XM
N/A
Ready
Last CBC XM batch was out of limits and current CBC
XM batch is within limits.
XM
N/A
Analytical Station
The last XM batch is outside limits.
stopped if XM
AutoStop option (1
batch out) is enabled.
System Status/Recovery
Corrective Action
1. Ensure all results are reportable. If necessary, exclude non-reportable
results from the statistics.
2. Check the results for a nonrandom sample distribution.
r If a nonrandom sample distribution is the problem, document the
problem based on your laboratory's protocol and continue processing
samples.
r Otherwise, troubleshoot for a possible Analytical Station problem.
1. Ensure all results are reportable. If necessary, exclude nonreportable results
from the statistics.
2. Check the results for a nonrandom sample distribution.
r If a nonrandom sample distribution is the problem, document the
problem based on your laboratory's protocol and continue processing
samples.
r Otherwise, troubleshoot for a possible Analytical Station problem.
Status only: no action necessary.
1. Check the results for a nonrandom sample distribution.
r If a nonrandom sample distribution is the problem, document the problem based
on your laboratory's protocol and continue processing samples.
r Otherwise, troubleshoot for a possible Analytical Station problem.
Status only: no action necessary.
1. Check the results for a nonrandom sample distribution.
r If a nonrandom sample distribution is the problem, document the problem based
on your laboratory's protocol and continue processing samples.
r Otherwise, troubleshoot for a possible Analytical Station problem.