6.5
WORKSTATION MESSAGES
Table 6.17 Workstation Messages - A through B
Workstation
Log Message and Icon on Command Center
<CBC / Diff / Retic / Retic Calc> XM all batches were deleted.
<CBC / Diff / Retic / Retic Calc> XM last batch was deleted.
<CBC / Diff / Retic / Retic Calc> XM targets and/or limits were
automatically updated.
A Control Lot number archive file was created.
AMC Timed out waiting for condition %1.
An archive file with patient samples was created
An unexpected error has occurred while processing sample <Sample
ID>, <Cass/Pos>
Auto numbering has been enabled. Starting auto number is
<Auto-Numbering Start ID>.
Automatic CBC calibration factors updated from Analyzer.
Automatic CBC calibration factors updated from Analyzer.
Automatic CBC calibration factors updated from GEN•S (WBC=
<9.999>, RBC=<9.999>, HGB=<9.999>, MCV=<9.999>,
PLT=<9.999>, MPV=<9.999>).
Auto-Numbering has been disabled.
Autostop has been triggered for Analytical Station
<Instrument#>.
PN 773023AC
Applicable
Log
Sample Results
System Status
Probable Cause
XM
N/A
Ready
The operator cleared all XM batches.
XM
N/A
Ready
The operator deleted the last XM batch.
XM
N/A
Ready
One or more XM targets and or limits of the specified group
were configured to automatically update upon batch
completion.
Control
N/A
Ready
Operator created a control archive file.
Event
N/A
Analytical Station
Communications error between Workstation and Analyzer.
stopped.
Event
N/A
Ready
Operator created an archive with a patient sample.
Event
Sample will not be
N/A
1. Workstation encountered a condition that causes a software
stored
2. Communications error between Workstation and Analyzer.
Event
SID of next run
Ready
Operator enabled the SID Auto-Numbering feature.
received will be
replaced by the
current Auto
Numbered SID.
Calibration
N/A
Ready
Received new automatic CBC calibration factors from
Analyzer.
Calibration
N/A
Ready
Workstation received new automatic CBC calibration factors
from Analyzer.
Calibration
N/A
Ready
Operator-initiated transmission of automatic CBC calibration
factors succeeded.
Event
N/A
Ready
Operator disabled the SID Auto-Numbering feature.
Event
N/A
Analytical Station
Autostop was triggered for one of the following reasons:
stopped
ToDo, Control, XB, or Data/Time error.
System Status/Recovery
exception, for example, when memory is overwritten.
TROUBLESHOOTING
WORKSTATION MESSAGES
Corrective Action
Status only; no action necessary.
Status only; no action necessary.
Status only; no action necessary.
Status only; no action necessary.
1. Acknowledge the error and reset the Analytical Station
2. If message persists, call your Beckman Coulter Representative.
Status only: no action necessary.
1. Restart the Workstation.
2. If the problem persists, reset the Analyzer.
3. If the problem persists after Analyzer reset, call your Beckman Coulter
Representative.
Status only; no action necessary.
Status only; no action necessary.
Status only: no action necessary.
Status only; no action necessary.
Status only; no action necessary.
Acknowledge error to clear Autostop condition.
6
6-49

Hide quick links: