Advertisement

Table 6.12 System Messages - R (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Run Aborted. PC Not
SYSTEM FAULT
N/A/
Ready.
and
NOT READY
R/W Data Mem.
NOT READY
N/A
Interference
PN 773023AC
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
No
Instrument Log
N/A
AMC Reported Error - Run
Aborted, PC Not Ready
Yes
Instrument Log
Data not transmitted.
Analyzer memory
read/write failure. Reset
Analytical Station.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
AMC to Workstation
communication error. This
could indicate:
r The Workstation is OFF.
r The cables connecting the
Analyzer and the
Workstation are loose or
disconnected.
r A hardware or software
malfunction.
Stopped - inoperable.
Internal memory error.
TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Corrective Action
1. Ensure the Workstation is on. If not, turn on the
Workstation, log onto the Workstation and go to
step 4.
2. Ensure the cables between the Analyzer and the
Workstation are connected. If not, connect the
cables and go to step 4.
3. If the Workstation is turned on, shutdown and
restart it, then go to step 4.
4. Reset the Analytical Station.
5. Continue normal operation. If message persists,
call your Beckman Coulter Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
6-41
6