Beckman Coulter LH 780 Instructions Manual page 185

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

Table 6.5 System Messages - D (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Dil1 Card -15V Error
(current status)
DIL1 CARD -15V
ERROR
Dil4 Card 12V Error
(current status)
DIL4 CARD
12V ERROR
Diluter Board Not
(current status)
N/A
Answering
Diluter Card 1 Failure
NOT READY
DILUTER CARD 1
FAILURE
Diluter Card 2 Failure
NOT READY
DILUTER CARD 2
FAILURE
Diluter Card 3 Failure
NOT READY
DILUTER CARD 3
FAILURE
Diluter Comm. Error
NOT READY
N/A
PN 773023AC
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
-15 Volts on Diluter 1
Alarm" message printed in
Card outside limit. Power
footer of sample report
down/up Analytical
(depends upon when error
Station.
occurred).
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
12 Volts on Diluter 4 Card
Alarm" message printed in
outside limit. Power
footer of sample report.
down/up Analytical
Station.
No
Instrument Log
Data not transmitted.
Analyzer-to-Diluter
communication failure.
Reset Analytical Station.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Diluter 1 I/O card failure
occurred).
(<Tube ID>, <Cass/Pos>).
Reset Analytical Station.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Diluter 2 I/O card failure
occurred).
(<Tube ID>, <Cass/Pos>).
Reset Analytical Station.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Diluter 3 I/O card failure
occurred).
(<Tube ID>, <Cass/Pos>).
Reset Analytical Station.
No
Instrument Log
Data displayed if available
(depends upon when error
Diluter communication
occurred).
failure. Reset Analytical
Station.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
U35 on the Diluter 1 card failed
(outside ±8% limit).
If this error occurs it will
not halt the action in
process, but it will prevent
initiation of a new sample,
unless the error condition
is corrected.
The Analytical Station
U5 on the Diluter 4 card failed
(outside ±5% limit).
stops processing samples.
If this error occurs, it will
not halt the action in
process but will prevent
initiation of a new sample
until the error condition is
corrected.
Stopped - inoperable.
Analyzer-to-Diluter
communication error.
Stopped - inoperable.
Diluter 1 card not answering.
Stopped - inoperable.
Diluter 2 card not answering.
Stopped - inoperable.
Diluter 3 card not answering.
Stopped - inoperable.
Failure to communicate
commands to Diluter.
TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Corrective Action
1. Power down the Analytical Station:
a. On the Numeric Keypad, press POWER OFF.
b. At the Power Supply, turn off the MAIN POWER
switch.
2. Power up the Analytical Station:
a. Turn on the MAIN POWER switch.
b. Press POWER ON.
3. If the message persists, call your Beckman Coulter
Representative.
1. Power down the Analytical Station:
a. On the Numeric Keypad, press POWER OFF.
b. At the Power Supply, turn off the MAIN POWER
switch.
2. Power up the Analytical Station:
a. Turn on the MAIN POWER switch.
b. Press POWER ON.
3. If the message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If the message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If the message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If the message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If the message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If the message persists, call your Beckman Coulter
Representative.
6-21